BS EN IEC 61453 BS EN 61453 Ed.3.0 Nuclear instrumentation - Scintillation gamma ray detector systems for the assay of radionuclides - Calibration and routine tests

Source:
IEC
Committee:
NCE/2 - Radiation protection and measurement
Categories:
Nuclear energy. General
Comment period start date:
Comment period end date:
Number of comments:
0

Comment by:

Scope

This document specifies methods of calibration and routine tests of scintillation detector systems for the measurement of gamma-ray energies and emission rates of radionuclides and the assay of radioactivity.

This document is applicable to scintillation detector systems based on inorganic scintillators for photon measurements.

Typical applications include radionuclide identification and assay in various industrial, environmental, and medical applications. The detector system consists of three major components: a scintillating material that produces photons of light when ionizing radiation interacts with it; one or more photo sensor, such as photomultipliers, photodiodes or other type of sensors (e.g., Si-photomultiplier), optically coupled to the scintillator, which convert the light photons to an amplified electrical pulse or pulses; and associated electronics which powers the photo sensor and processes the output signal.

Both energy calibration and efficiency calibration are covered in this document. The following three techniques are considered:

a)      total spectrum counting (see 4.1) which employs a system that counts all pulses above a low-energy threshold (see 6.1, 6.2 and 6.3);

b)      single-channel analyzer (SCA) counting (see 4.2) which employs a system with a counting channel established through upper and lower energy boundaries (see 6.1, 6.2, and 6.3);

c)      multichannel analyzer (MCA) counting (see 4.3) which employs a system in which multiple counting windows are utilized. This technique allows measurements for which the continuum under the total absorption peak may be subtracted without introducing unacceptable error. In case of overlapping peaks in the spectrum, a multichannel analyzer with access to a peak deconvolution program is necessary. This case is not covered by this standard.

Read draft and comment

Comment on proposal

Required form fields are indicated by an asterisk (*) character.


Please email further comments to: debbie.stead@bsigroup.com

Follow standard

You are now following this standard. Weekly digest emails will be sent to update you on the following activities:

You can manage your follow preferences from your Account. Please check your mailbox junk folder if you don't receive the weekly email.

Unfollow standard

You have successfully unsubscribed from weekly updates for this standard.

Error