If you have difficulty in submitting comments on draft standards you can use a commenting template and email it to admin.start@bsigroup.com. The commenting template can be found here.
This document applies to medical electron accelerators when used, for therapy purposes, in human medical practice.
This document applies to medical electron accelerators that deliver a radiation beam produced by accelerated electrons, either as electron radiation (specified by nominal energy, 1-50 MeV) or as X-radiation generated from electrons (specified by nominal accelerating potential, 1-50 MV) with an achievableabsorbed dose rates between 0,001 Gy s−1 and 1 Gy s−1 at the ERP equipment reference point from the radiation source and at reference treatment distances between 0,4 m and 2 m from the radiation source.
NOTE 1 This document applies to conventional dose-rates up to 1 Gy·s⁻¹. Ultra-high dose-rate (FLASH) operation is outside the scope and shall be characterized separately by the manufacturer.
NOTE 2 Manufacturers providing Ultra high dose-rate (FLASH) capable modes shall declare functional performance characteristics separately, including beam stability, dose per pulse, and dosimetry traceability
The present standard describes measurements and test procedures to be performed by the manufacturer at the design and construction stage of a medical electron accelerator; it does not prescribe acceptance tests to be performed after installation at the purchaser's site, but test procedures have been chosen to align with those appropriate for acceptance tests.
This document specifies test procedures for the determination and disclosure of functional performance characteristics, knowledge of which is deemed necessary for proper application and use of a medical electron accelerator and that are to be declared in the accompanying documents together with the deviation or variation from the declared values to be expected under specific conditions in normal use. A format for presentation of functional performance values is given in Annex A.
It is recognized that uncertainties in the test methods must be allowed for when assessing performance. As far as possible test methods described do not focus on the measurement errors but the equipment deviations, this in the expectation that more accurate test methods will be developed over time.
It is not intended that this document should in any way inhibit the future development of new designs of equipment that can have operating modes and parameters different from those described herein, provided that such equipment achieves equivalent levels of performance for the treatment of patients.
Except where otherwise stated, this document assumes that the medical electron accelerators has an isocentric gantry. Where the equipment is non-isocentric, the description of performance and test methods can be suitably adapted; any such adaptation shall be described by the manufacturer .
You are now following this standard. Weekly digest emails will be sent to update you on the following activities:
You can manage your follow preferences from your Account. Please check your mailbox junk folder if you don't receive the weekly email.
You have successfully unsubscribed from weekly updates for this standard.
Comment on proposal
Required form fields are indicated by an asterisk (*) character.