This document is limited to Luer activated needle-free connectors (LANCs) for intravascular applications, which open and permit access to the fluid conduit when a male Luer connector is inserted.
This document applies only to the valve end of LANCs.
This document applies to LANCs incorporated into a medical device.
This document does not address valves that are opened by a female Luer connector.
This document does not address the connection features of the non-valve end of LANCs and/or other integrated features (e.g. stopcock handles) (Figure 1).
Please see draft for Figure 1 — Examples of a finished medical device (left) or subassemblies (middle & right) incorporating LANCs
For LANCs that are incorporated into a finished device, such as closed system transfer devices (CSTD), additional requirements may need to be considered.