BS ISO 24645 ISO 24645 General requirements for Luer activated needle-free connectors (LANCs) for intravascular applications

Source:
ISO
Committee:
CH/212 - IVDs
Categories:
Syringes, needles and catheters
Comment period start date:
Comment period end date:
Number of comments:
0

Comment by:

Scope

NOTE See A.3.2.

This document is limited to Luer activated needle-free connectors (LANCs) for intravascular applications, which open and permit access to the fluid conduit when a male Luer connector is inserted.

This document applies only to the valve end of LANCs.

This document applies to LANCs incorporated into a medical device.

This document does not address valves that are opened by a female Luer connector.

This document does not address the connection features of the non-valve end of LANCs and/or other integrated features (e.g. stopcock handles) (Figure 1).

Please see draft for Figure 1 — Examples of a finished medical device (left) or subassemblies (middle & right) incorporating LANCs

For LANCs that are incorporated into a finished device, such as closed system transfer devices (CSTD), additional requirements may need to be considered.

Read draft and comment

Comment on proposal

Required form fields are indicated by an asterisk (*) character.


Please email further comments to: debbie.stead@bsigroup.com

Follow standard

You are now following this standard. Weekly digest emails will be sent to update you on the following activities:

You can manage your follow preferences from your Account. Please check your mailbox junk folder if you don't receive the weekly email.

Unfollow standard

You have successfully unsubscribed from weekly updates for this standard.

Error