BS EN 1422 EN 1422 Sterilizers for medical purposes. Ethylene oxide sterilizers. Requirements and test methods

Source:
CEN
Committee:
CH/198 - Sterilization and Associated Equipment and Processes
Categories:
Sterilizing equipment
Comment period start date:
Comment period end date:
Number of comments:
0

Comment by:

Scope

1.1 This document specifies the requirements and the relevant tests for automatically controlled sterilizers employing ethylene oxide (EO) gas as the sterilant, either as a pure gas or a mixture with other gases, being used for the sterilization of medical devices and their accessories.

1.2 This document specifies requirements for ethylene oxide sterilizers (EO sterilizers) for:

— the performance and design of sterilizers intended to deliver a process capable of sterilizing medical devices;

— the equipment and controls of these sterilizers, needed for operation, control and monitoring of the sterilization processes;

— the test equipment and test procedures used to verify the sterilizer performance specified by this document.

1.3 This document does not cover sterilizers which employ the injection of EO or mixtures containing EO directly into packages or into a flexible chamber.

Read draft and comment

Comment on proposal

Required form fields are indicated by an asterisk (*) character.


Please email further comments to: debbie.stead@bsigroup.com

Follow standard

You are now following this standard. Weekly digest emails will be sent to update you on the following activities:

You can manage your follow preferences from your Account. Please check your mailbox junk folder if you don't receive the weekly email.

Unfollow standard

You have successfully unsubscribed from weekly updates for this standard.

Error