BS IEC 61094-9 Ed.1.0 Electroacoustics - Measurement microphones. - Part 9: Specifications for transfer standard microphones

Source:
IEC
Committee:
EPL/29 - Electroacoustics
Categories:
Electroacoustics | Accessories
Number of comments:
0

Scope

This part of IEC 61094 specifies the geometrical configuration and essential electroacoustic characteristics for condenser microphones used as transfer standards for the dissemination of the unit of sound pressure and for sound pressure measurements of high accuracy.

The specifications are intended to optimise reproducibility in pressure sensitivity calibration by comparison in couplers and when coupled to a given sound calibrator, in the frequency range from 10 Hz to 1 260 Hz, when using the transfer standard microphone as a reference device.

The specifications do not ensure that a transfer standard microphone can be used at frequencies above 1 260 Hz. An LS2P reference microphone is needed at frequencies above 1 260 Hz. Specifications for transfer standard microphones at higher frequencies are included because they are necessary for the intended use.

This part of IEC 61094 also supplements the mnemonic system established in IEC 61094-1 for classifying measurement microphones, to include transfer standard microphones.

Comment on proposal

Required form fields are indicated by an asterisk (*) character.


Please email further comments to: debbie.stead@bsigroup.com

Follow standard

You are now following this standard. Weekly digest emails will be sent to update you on the following activities:

You can manage your follow preferences from your Account. Please check your mailbox junk folder if you don't receive the weekly email.

Unfollow standard

You have successfully unsubscribed from weekly updates for this standard.

Error