Scope
1.1.1 Scope
Replacement:
This Part 2 is applicable to electrically powered LABORATORY CENTRIFUGES.
It is possible that all or part of the equipment falls within the scope of one or more other Part 2 standards of IEC 61010 as well as within the scope of this standard . In that case, the requirements of those other Part 2 standards will also apply
This document is not applicable to electrically powered LABORATORY CENTRIFUGES. manufactured before the date of its publication.
1.1.2 Equipment excluded from scope
Addition:
Add the following new item:
aa) IEC 60034 (Rotating electrical machinery);
1.2 Object
1.2.1 Aspects included in scope
Addition:
Add the following new items:
aa) contact with moving parts (see 7.3);
bb) LABORATORY CENTRIFUGE movement during any DISRUPTION (see 7.3.101);
cc) high energy chemical reaction after ROTOR DISRUPTION (see 7.7.2.2 l));
dd) ineffectiveness of BIOSEALS (see 13.101)
1.2.2 Aspects excluded from scope
Addition:
Add the following new items:
aa) additional precautions which may need to be observed when centrifuging materials 150 which are flammable or explosive (see 5.4.101);
bb) additional precautions which may need to be observed when centrifuging materials that could react chemically with sufficient vigour to cause a HAZARD (see 5.4.101).
1.4 Environmental conditions
1.4.1 Normal environmental conditions
Replacement:
Replace item c) by the following:
c) temperature 2 °C to 40 °C; 158
1.4.2 Extended environmental conditions
Replacement:
Replace item c) by the following:
c) ambient temperatures below 2 °C or above 40 °C;