Find out what cookies we use and how to disable themThis clause of Part 1 is applicable except as follows:1.1.1 Equipment included in scope
1.1.1 Equipment included in scope
Replacement:Replace the second paragraph by the following:
This part of IEC 61010 specifies safety requirements for electrical equipment and its accessories within the categories a) through c), wherever it is intended to be used, whenever that equipment incorporates one or more of the following characteristics:
– A refrigerating system that is acted on or impacted by an integral heating function such that the combined heating and refrigerating system generates additional and/or more severe hazards than those for the two systems if treated separately, and when containing or intended for use with hazardous fluids, whether flammable or under high pressure, such that additional and/or more severe hazards resulted than those for non-hazardous fluids.
– The materials being treated in the intended application introduce significant heat into the refrigerating system, so that the refrigerating system in the application yields additional and/or more severe hazards than those for the refrigerating system if operated at the maximum rated ambient temperature alone.
– An irradiation function for the materials being treated presenting additional hazards.
– A function to expose the materials being treated to damp heat, carbon dioxide, salt mist, or other substances which can result in additional hazards.
– A function of mechanical movement presenting additional hazards.
– Provision for an operator to walk in to the operating area to load or unload the materials being treated.
Addition:
Add the following text after the last paragraph:
NOTE 101 Examples of such equipment include environmental testing and plant growth test chambers, refrigerating circulators which incorporate heating, and recirculating coolers for extracting heat.
It is possible that all or part of the equipment falls within the scope of one or more other Part 2 standards of IEC 61010 as well as within the scope of this standard. In that case, the requirements of those other Part 2 standards also apply. This document is intended for application when one or more of the additional hazards described in the above dashed listed items are introduced. However, when the equipment incorporates only a refrigerating system or only a heating function or a combination of the two without introducing the additional hazards described in the above list, then IEC 61010-2-011 or IEC 61010-2-010 or both, as appropriate, apply instead of this Part 2-012.
See further information in the flow chart (Figure 102) for selection process and guidance in the Introduction.
NOTE 102 Subclause 3.1.107and Annex BB provide the definition and requirements for the protection ofpeople who are inside walk-in equipment.1.1.2 Equipment excluded from scope
1.1.2 Equipment excluded from scope
Addition:
Add the following items after item j):
aa) equipment for the heating, cooling, and ventilation of laboratories;
bb) sterilizing equipment.
1.2 Object
1.2.1 Aspects included in scope
Addition:
Add the following items after item g):
aa) biohazards(see 13.101);
bb) hazardous chemical substances(see13.102).
You are now following this standard. Weekly digest emails will be sent to update you on the following activities:
You can manage your follow preferences from your Account. Please check your mailbox junk folder if you don't receive the weekly email.
You have successfully unsubscribed from weekly updates for this standard.
Comment on proposal
Required form fields are indicated by an asterisk (*) character.